A free-standing slimline console
design providing an ISO-5/Class 100 atmosphere
in an operator workstation environment.
Completely self-contained, it functions
effectively in an uncontrolled area.
Individually tested and certified
to meet or exceed the specifications
for Laminar Flow Clean-Air Devices
as per Institute of Environmental
Science IES-RP-CC-002-86.
Specifications (see details)
Filter - Three inch HEPA (High Efficiency Particulate Air) type, aerosol challenged,
zero probed. Aluminum frame minipleat design, Minimum efficiency of 99.99%
at .3 microns.
Prefilter - 1" bonded polyfiber media minimizes
surface-loading. Quick-release aluminum
frame. Easily accessible from front
of unit. 78 % arrestance 56-57 ASHRAE.
Airflow - Average measured clean air velocity
90 fpm (+/-10 fmp). All measured values
within +/- 20% uniformity.
Construction - Woodboard construction for minimum
noise & vibration. White and blue
plastic laminate finish. Clear acrylic
side panels for visibility. Rear-loaded
HEPA filter. Protective plastic egg crate diffuser screen in front of HEPA filter
keeps uniform air velocity within
work area.
Electrical & Power Requirements
- All electrical components are UL Approved. Standard 115V, 15 Amp, 1 pH,
60 Hz. SJO power cord w/ 3-prong ground plug.
NEC Electrical Code
Motor/Blower Assembly - High efficiency motors to reduce electrical load. Motors are thermally
protected & self-lubricating. High capacity metal forward curved blowers
with direct drive motors. Fan statically balanced to maintain constant
airflow. Motors/Blowers are selected to maintain constant required airflow
rate up to 2x the initial static pressure.
Lighting - Recessed 2 pin fluorescent lighting. 125 ft. Candle +/- 20 ft. Candles.
Plastic prismatic lens cover maximizes light diffusion.
Controls - Solid state, adjustable airflow control,
80-120 fpm, 15A max. load capacity,
Internally mounted to prohibit accidental
change. Illuminated switches on face
panel for motor & lights.
|